Model #: nghxr

eXact

Spectrophotometer

Offered in both Standard and Advanced models, the eXact combines industry-leading functionality with an easy-to-use interface to deliver the best solution for controlling, managing, and communicating color across the entire color workflow.




Specifications

Aperture 1.5 mm, 2 mm, 4 mm or 6 mm
Battery

Lithium Ion, 7.4VDC, 2200mAh

BestMatch Feature Yes
Calibration Automatic on integrated white reference
Color Spaces CIE L*a*b*, CIE L*C*h°, CIE XYZ and Yxy
Color Differences CIE ∆E* (1976), ∆ECMC, CIE ∆E* (2000), and CIE ∆E* (1994)
Communication Interface USB 2.0; Bluetooth (optional) - Class II
Configuration Options Bluetooth, NetProfiler
Density Range 0.0 D – 3.0 D
Density Standards Status A, E, I, T and G
Dimensions (length, width, height) 18 cm x 7.8 cm x 7.6 cm
7.1" x 3.1" x  3.0"
Embedded NetProfiler Support Yes
G7 Certification Yes with embedded G7, PSO, JapanColor job templates
Humidity 30 to 85% RH (non-condensing)
Illuminants A, C, D50, D55, D65, D75, F2, F7, F11, D12
Instrument Display Color touch-screen, 320 x 240 pixels
Inter-Instrument Agreement Average: 0.25 ΔEab,
Max: 0.45 ΔEab (for M3: 0.55 ΔEab)
(Measurements using X-Rite manufacturing standards at a temperature of 23°C +/- 1°C, 40-60% RH for all measurement modes on 12 BCRA color tiles and a white ceramic reference (D50, 2°))
Languages Supported Chinese Simplified, Chinese Traditional, Czech, Danish, Dutch, English, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Polish, Portuguese, Russian, Slovenian, Spanish, Swedish, Turkish, Vietnamese
Light Source Gas filled tungsten (illuminant type A) and UV LED
Industry Standards ISO 5-4:2009(E), ISO 13655:2009
Measurement Conditions M0, M1 (part 1), M1 (part 2) , M2, M3; also available in Xp variant for flexible packaging substrates
Measurement Geometry 45°/0° ring illumination optics, ISO 5-4:2009(E)
Measurement Time < 1 sec
NetProfiler Support Yes
Observers 2° and 10°
Operating Temperature Range 10°C to 35°C
50°F to 95°F
PantoneLIVE Support Yes; Embedded Pantone Libraries
Pass / Fail Yes
Sample/Measurement Max. 5,000 samples
Short Term Repeatability - White 0.05 ΔEab (standard deviation) White BCRA (Error compared to mean value of 10 measurements every 5 seconds)
Short Term Repeatability - Density '+/- 0.01 D for CMYK (Status E or Status T measurements, maximum error compared to mean value of 10 measurements every 5 seconds at 2.0 D except for M3 Yellow at 1.7 D)
Software Development Kit

Yes

Software Modes Densitometer, Standard, Advanced
Spectral Analyzer DRS spectral engine
Spectral Interval 10 nm
Spectral Range 400 nm to 700 nm
Spectral Reporting 400 nm to 700 nm with 10 nm interval
Storage Temperature Range -20° to 50°C
-4° to 122°F
Standards More than 20'000 standards (incl. data for all M modes); New Color Library Tool to Create and Store Standards
Voltage 12 VDC @ 2.5 A
Warranty 1 year
Weight  700 g
1.6 lbs.
What's Included? USB cable, power supply, CD with software and documentation, docking/charging station, quick start guide, carrying case, device certificate
3rd Party Software Supported

X-Rite ColorCert, X-Rite Color iQC, X-Rite InkFormulation, X-Rite NetProfiler and many others


Need a Quote? Contact Sales(888) 800-9580