Fin du support de Windows 10 : Découvrez comment cela impacte notre logiciel - Windows 10 EOL


Printer QA - Higher DeltaE than specified 'PASS' in the report

Question: I generated a QA report based on the UGRA/FOGRA Media Wedge verification according to ISO 12647-7 tolerances for proof printing The Max Delta E for overall colors resulted in a value of of 6.28 and it has 'PASS'ed  in the report. However the threshold is Delta E = 6. Is this a bug in the report generation?

Answer: No, this is not a bug. The Delta E tolerances in the 12647-7 specification are single digits (e.g. 6, not 6.0). So, every value below 6.5 will 'PASS' in the verification report. This is also true for the other tolerances.
 


Pour un devis, contactez le service commercial ou appelez au +33 6 74 63 33 79.

Pour du support, contactez le support ou appelez au 00 800 700 300 01.