Resetting the Lamp Meter on SpectraLight III

The SPL III allows the user to replace lamps.  After lamp replacement, you should reset the time meter for the new lamp(s).  To reset the lamp time meter, proceed as follows:

StepAction
1.Press and hold the source button for the replaced lamp for five seconds
2.The display will show "PRESS SET/SAVE TO CONFIRM RESET"
3.Press the Set/Save button
4.The hours are set to zero (0)