Fine del supporto per Windows 10: scopri come questo influirà sul nostro software - Windows 10 EOL


7000A "Black Level to High" when using SAV aperture.

Several things can be done to try to resolve this issue.
1) Clean the lens and mirror make sure the aperture plate is snug and test.
2) Remove excess barium from SAV aperture with a dry q-tip swab and test.
3) Set the lense to VSA setting and calibrate with the SAV aperture.

Note: If there is still an issue there may be an allignment issue. In this case it will have to come in for a re-allignment.


Per preventivi contatta l’ufficio commerciale o il +39 347 65 26 778. Per l’assistenza tecnica contatta il supporto o il 00 800 700 300 01 (da linea fissa).