If the MA 58 instrument reports the error message "Maximum Zero Reflectance Exceeded", the black glass calibration tile must be thoroughly cleaned. Thoroughly clean the black reference tile with a damp (water only) lint free cloth and then dry thoroughly with a dry lint free cloth. If more aggressive cleaning is required, iospropyl alcohol may be used instead of water.