Manutenzione programmata: I nostri siti e servizi potrebbero subire interruzioni intermittenti il 16 agosto dalle 20:30 alle 22:30 (EDT). Grazie per la vostra pazienza.


Invalid Reading or Unrecognizable strip error

Getting and Invalid reading or unrecognizable strip error.
Invalid Reading or Unrecognizable Strip Error

These are messages you will get when your unit won't recognize or read a control strip.

 Possible Causes:

  • Wrong Control strip selected, check code being used.
  • Not enough Leader, strips must have more than 1.25 inches before the first target.
  • Calibration, re-calibrate  With a current calibration strip (less than 18months old),
  • Was the strip inserted in the right direction?  Some strips are not bi-directional.
  • Was the strip aligned with the center diamond? and did it run straight all the way through?
  • Lamp strength? (replacement level is anything below 93%)
  • Motor speed


Domande? Vi serve un preventivo? Contattateci(888) 800-9580