Resetting an X-Rite MA68

Under some conditions it becomes necessary to reset the microprocessor in an X-Rite Multi-Angle spectrophotometer. Be aware however that an instrument reset is a 'last-resort' decision. Any stored standards or sample data will be deleted if you continue! The following procedure will return your MA (multi-angle) unit factory defaults:

The multi-angle instrument will then display "MEMORY INITIALIZED BY OPERATOR Press Any Key".

The unit is now returned to factory defaults. Immediately following this procedure you must continue by calibrating the unit before measurements may be taken (see Related Support).

Calibrating the MA58 & MA68