Fin del soporte técnico para Windows 10: Descubra cómo afecta esto a nuestro software - Windows 10 EOL


IT8.7-2 Scan Targets cannot be measured with iO Table

Question: iO table does not measure my IT8.7-2 Scan Target correctly when selecting the 'IT8 Kodak Format.txt' (or Agfa or Fuji) reference file in MeasureTool 5. Used 'Patch' measurement mode, but iO measures the patches in wrong order. How can I solve this?

Answer: The IT8.7-2 Scan Target Reference Files 'IT8 Agfa Format.txt', 'IT8 Fuji Format.txt' and 'IT8Kodak Format.txt' are not usable for the iO Table. They are designed for SpectroScan and refer to hard-coded driving commands for the SpectroScan.

The IT8 Scan Targets can be measured manually in patch mode with the Eye-One by using the above mentioned reference files.



¿Tiene preguntas? ¿Necesita un presupuesto? Contáctenos.  +34 662 96 70 49

¿Tiene preguntas tecnicas?  Contáctenos.  00 800 700 300 01