Make sure white calibration standard and black trap are clean.
Press "menu" (key #1) twice to enter "MENU PAGE 2". (see note)
NOTE: If "read" is displayed above key #1 then"double click" the key as you would a mouse button to enter "MENU PAGE 2"
Press "cal" (key #2) to enter calibration function.The software datecode and copyright are momentarily displayed then ''READ WHITE CALIBRATION PLAQUE'' is displayed.
Position the instrument on the white standard and take measurement.The following sequence is displayed:
Place the instrument over the black trap.
Lower the instrument to the target window to take measurement.