Calibrating the ATD / ATS / ATS Pub System

Calibrating the ATD / ATS / ATS Pub Head

Calibration is an essential part of the maintenance procedure that helps insure accurate and repeatable results from your scanner.  To calibrate the ATS or ATD head, follow the steps below.

  1. Make sure the calibration plaques are clean.  The plaques are located under the head when it's docked.  To expose them, simply push the head away from the docking station and push the trap door toward the docking station.  You will see white and black cal plaques.  To clean the plaques, use a clean dry cotton cloth and simply polish the plaques.  Do not use any chemicals, oils, glass cleaner, or soap.  After the calibration plaques are clean, push the head back to the docking station.

  2. Click on the 'Instrument' drop down menu.

  3. Click on 'Scanning' and then click on 'Calibrate'.

The calibration procedure will take only twenty seconds or so......All done!

 

Calibrating the ATD / ATS / ATS Pub Track

Calibration is an essential part of the maintenance procedure that helps insure accurate and repeatable results from your scanner.  To calibrate the ATS or ATD track,  follow the steps below.

  1. Make sure the track is clean.  To clean the track, spray the surface with glass cleaner and whip down.  Follow up with a clean dry cotton cloth.  Do not use any chemicals, oils, or soap.  

  2. Click on the 'Instrument' drop down menu.

  3. Click on 'Scanning' and then click on 'Calibrate Track'.

The calibration procedure will take a few seconds or so......All done!

Need a Quote? Contact Sales(888) 800-9580

Technical Questions? Contact Support(888) 826-3042