There are 4 available apertures for the eXact handheld; NOTE: These are not changeable in the field.
1.5mm, 2.0mm, 4.0mm, 6.0mmThere are 4 available apertures for the eXact handheld; NOTE: These are not changeable in the field.
Aperture & target window size
For each eXact aperture size, the target window size is 1 mm (on radius) larger than the spot measurement area size. This 1 mm over illumination of the measurement area is required by ISO5/4. The white area around the patch, does not influence the measurement data, but does require the operator to center the measurement sample in the target window opening.
Available aperture sizes:
Aperture Size
Measurement area size
Opening in target window
Recommended patch size
Screening Range
1.5mm
1.5mm
3.5mm
2mm - 3mm
175 lines/inch or 69 lines/cm or above
2mm
2mm
4mm
3mm - 5mm
133 lines/inch or 52 lines/cm or above
4mm
4mm
6mm
5mm - 7mm
65 lines/inch or 26 lines/cm or above
6mm
6mm
8mm
7mm or larger
It is always recommended to use the largest aperture size possible.
The illustration below shows how this minimum measurement area appears in relation to the target window opening.
The black circle represents the target window opening.
The white circle represents the actual measurement area (which equals the aperture size).
The magenta square represents the minimum suggested patch size.