Application :WinDense Medical
Vendor :Photoware Limited
Platform :Win 3.1x.. NT 4
Version :v17
Cabling :SE108-92 (cable), SE108-99 (coaxial to phone jack - power supply adapter) and SE30-61 (power supply)
Notes :Per latest info from the folks at Windense: the most current Medical package is v17. Software operates on all versions of Microsoft Windows from v3.1x to Win NT v4.

 

WinDense Medical is a complete data receiver, storage system, and plotting utility for use with X-Rite strip reading densitometers. Per notes from their manual, the following modes must be set into the instrument:

I/O TYPE= CUSTOM
BAUD=9600
PIN5=CTS (Was: Off Edited 2/18/2000 jgm)
FORMAT =NEW
RCI=OFF
DEC PT=ON
CR=CRLF
COMP=OFF
XMIT=AUTO
WIDTH=FULL

In the WinDense software, user must configure the system for communications with the X-Rite instrument. A 'setup' button allows the selection of several different densitometers. Select the X-Rite 380-391 from this list. A graphic icon allows the edit of additional settings. To work with all X-Rite densitometers ensure the following software settings:

BAUD= 9600 
DATA BITS =8 
STOP BITS=1 
PARITY=NONE 

Finally, ensure that the user has selected the appropriate communications port.