Selecting a SpectraLight Daylight Source

Q - Which daylight source do I need to order on my Spectralight?

A - The Daylight color temperature that you choose should follow a specification given to you or in accordance to your industry.

- D75 North Sky Daylight at 7500K is used for visual evaluation of opaque materials as outlined by ASTM D1729.

- D65 Average North Sky Daylight at 6500K is used to provide visual correlation with spectrophotometric readings or in conformance with Automotive, Textile, European or Japanese standards.

- D50 Noon Sky Daylight at 5000K is used for the evaluation of color quality and uniformity in conformance with ANSI and ISO specifications governing the graphic arts.

Need a Quote? Contact Sales(888) 800-9580

Technical Questions? Contact Support(888) 826-3042