Epson C80 and the 390

Users will need to configure the 390/391 to communicate with thyeir C80 if desired.
Application : Epson C80
Vendor : Epson
Platform : Parallel Printer
Version : NA
Cabling : SE130-SERPAR (RJ45 to Centronics cable)
Notes : This Epson printer is a PARALLEL device! A serial to parallel converter must be used! The above cable satisfies this requirement. Note that the "SERPAR" converter operates at a fixed baud rate of 9,600!

If your densitometer has a "Stylus" I/O preset, select it now.  If "Stylus" is not an option use the "CANON1" setting. There is nothing to configure on the printer!  It will operate correctly (out-of-the-box).

Note: In December of 2001 it was found that several of the serial to parallel cables had been incorrectly wired.  This is evidenced by a "Check Printer" warning on the densitometer.  These cables will still function but enabling communications requires a single change in the densitometer.  After saving either "Stylus" or "CANON1", return to the configuration page and using Custom Presets, change "PIN5" to "BUSY".  (12/22/2000 jgm)


   The support number for Epson is... 800-922-8911 begin_of_the_skype_highlighting
   The WebSite for Epson is... www.epson.com

Questions? Need a Quote? Contact Us.  1.800.248.9748