DTP70 Autoscan Spectrophotometer (Discontinued)


Date Created: 7/22/2004   Date Modified: 7/23/2004

Cleaning the Calibration Reference

1. Position the reading head to the center of the measurement path using ToolCrib (see related support articles on the left).

2. Disconnect the power supply from the instrument.

3. Remove the front cover by pulling forward on the lower portion. After the lower portion of the cover is free, rotate it upwards to free the upper portion from the rear cover. Set the cover aside.



4. Clean the calibration reference using a cotton swab or lint free cloth. Wipe from left-to-right to minimize lint collection. If needed, a small quantity of water or alcohol can be used to clean the calibration reference. Do not use any other solvents or glass cleaner.



5. Reinstall the front cover by inserting the top portion into the rear cover and rotating downward. Push the lower portion of the front cover inward until it is flush with the bottom housing.

6. Reconnect power.





   